« Volver al listado

CVE-2026-7251

Estado: Pendiente de análisisCrítica (9.3)—

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

🎯 Técnicas ATT&CK

Cómo se explota esta vulnerabilidad y qué consigue el atacante, en el lenguaje de MITRE ATT&CK.

Acceso remoto de red (AV:N, PR:N) a servidor VNC con credencial hardcodeada (CWE-259). Impactos: autenticación sin credenciales válidas (T1078) y ejecución de comandos vía interfaz de control (T1059).

Inferido por nuestro agente de análisis a partir de la descripción oficial, el vector CVSS y la CWE, y comprobado por un supervisor. Puede contener errores.

🛡️ Mitigaciones ATT&CK que cubren estas técnicas

Tecnologías afectadas (1)

⚠ Inferidas por IA a partir de la descripción — NVD aún no ha analizado esta CVE; no son CPE verificados.

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2026-7251",
  "cveTags": [],
  "metrics": {
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2026-7251",
          "role": "CISA Coordinator",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "yes"
            },
            {
              "technicalImpact": "total"
            }
          ],
          "version": "2.0.3",
          "timestamp": "2026-05-26T18:32:05.373317Z"
        }
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "ics-cert@hq.dhs.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 9.8,
          "attackVector": "NETWORK",
          "baseSeverity": "CRITICAL",
          "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 3.9
      }
    ],
    "cvssMetricV40": [
      {
        "type": "Secondary",
        "source": "ics-cert@hq.dhs.gov",
        "cvssData": {
          "Safety": "NOT_DEFINED",
          "version": "4.0",
          "Recovery": "NOT_DEFINED",
          "baseScore": 9.3,
          "Automatable": "NOT_DEFINED",
          "attackVector": "NETWORK",
          "baseSeverity": "CRITICAL",
          "valueDensity": "NOT_DEFINED",
          "vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
          "exploitMaturity": "NOT_DEFINED",
          "providerUrgency": "NOT_DEFINED",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "attackRequirements": "NONE",
          "privilegesRequired": "NONE",
          "subIntegrityImpact": "NONE",
          "vulnIntegrityImpact": "HIGH",
          "integrityRequirement": "NOT_DEFINED",
          "modifiedAttackVector": "NOT_DEFINED",
          "subAvailabilityImpact": "NONE",
          "vulnAvailabilityImpact": "HIGH",
          "availabilityRequirement": "NOT_DEFINED",
          "modifiedUserInteraction": "NOT_DEFINED",
          "modifiedAttackComplexity": "NOT_DEFINED",
          "subConfidentialityImpact": "NONE",
          "vulnConfidentialityImpact": "HIGH",
          "confidentialityRequirement": "NOT_DEFINED",
          "modifiedAttackRequirements": "NOT_DEFINED",
          "modifiedPrivilegesRequired": "NOT_DEFINED",
          "modifiedSubIntegrityImpact": "NOT_DEFINED",
          "modifiedVulnIntegrityImpact": "NOT_DEFINED",
          "vulnerabilityResponseEffort": "NOT_DEFINED",
          "modifiedSubAvailabilityImpact": "NOT_DEFINED",
          "modifiedVulnAvailabilityImpact": "NOT_DEFINED",
          "modifiedSubConfidentialityImpact": "NOT_DEFINED",
          "modifiedVulnConfidentialityImpact": "NOT_DEFINED"
        }
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "Eppendorf",
          "product": "BioFlo 320",
          "versions": [
            {
              "status": "affected",
              "version": "All",
              "versionType": "custom"
            }
          ],
          "defaultStatus": "unaffected"
        }
      ]
    }
  ],
  "published": "2026-05-26T18:16:55.347",
  "references": [
    {
      "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-146-01.json",
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-146-01",
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://www.eppendorf.com/software-downloads",
      "source": "ics-cert@hq.dhs.gov"
    }
  ],
  "vulnStatus": "Awaiting Analysis",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-259"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted."
    },
    {
      "lang": "es",
      "value": "Eppendorf BioFlo 320 es vulnerable debido a que el servidor VNC utiliza una contraseña codificada de forma rígida. Si un atacante remoto conoce la dirección de red de cualquier modelo BioFlo 320 con acceso remoto habilitado, puede obtener control total de la interfaz de usuario utilizando esta contraseña. Una vez conectado, el atacante tendría acceso total a todas las funciones del panel de control del BioFlo 320. El tráfico VNC no está cifrado."
    }
  ],
  "lastModified": "2026-07-24T08:10:00.150",
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}