« Volver al listado

CVE-2025-54304

Estado: AnalizadaCrítica (9.8)—

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network.

Leer descripción completaMostrar menos

The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

🎯 Técnicas ATT&CK

Cómo se explota esta vulnerabilidad y qué consigue el atacante, en el lenguaje de MITRE ATT&CK.

X11 expuesto en puerto 6000 accesible sin autenticación desde la red (AV:N/PR:N/UI:N) permite ejecución remota de código y escalada a root mediante matchbox-desktop.

Inferido por nuestro agente de análisis a partir de la descripción oficial, el vector CVSS y la CWE, y comprobado por un supervisor. Puede contener errores.

🛡️ Mitigaciones ATT&CK que cubren estas técnicas

Tecnologías afectadas (1)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2025-54304",
  "cveTags": [],
  "metrics": {
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2025-54304",
          "role": "CISA Coordinator",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "yes"
            },
            {
              "technicalImpact": "total"
            }
          ],
          "version": "2.0.3",
          "timestamp": "2025-12-05T20:03:18.970227Z"
        }
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 9.8,
          "attackVector": "NETWORK",
          "baseSeverity": "CRITICAL",
          "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 3.9
      }
    ]
  },
  "affected": [
    {
      "source": "cve@mitre.org",
      "affectedData": [
        {
          "vendor": "n/a",
          "product": "n/a",
          "versions": [
            {
              "status": "affected",
              "version": "n/a"
            }
          ]
        }
      ]
    }
  ],
  "published": "2025-12-04T15:15:58.773",
  "references": [
    {
      "url": "https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf",
      "tags": [
        "Product"
      ],
      "source": "cve@mitre.org"
    },
    {
      "url": "https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdf",
      "tags": [
        "Vendor Advisory"
      ],
      "source": "cve@mitre.org"
    },
    {
      "url": "https://www.thermofisher.com/order/catalog/product/4474779",
      "tags": [
        "Product"
      ],
      "source": "cve@mitre.org"
    }
  ],
  "vulnStatus": "Analyzed",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
      "description": [
        {
          "lang": "en",
          "value": "CWE-200"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer."
    }
  ],
  "lastModified": "2026-06-17T09:39:49.017",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware:-:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "3ADB7EF9-E6BF-468B-8B96-98C79409E924"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:thermofisher:ion_torrent_onetouch_2:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "6CD40B87-870A-4B96-8864-4687DA5AB439"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "sourceIdentifier": "cve@mitre.org"
}