CVE-2024-48971
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- Puntuación base: 9.3
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.22%
- Percentil entre todas las CVEs puntuadas: 11
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
🎯 Técnicas ATT&CK
Cómo se explota esta vulnerabilidad y qué consigue el atacante, en el lenguaje de MITRE ATT&CK.
- Explotación
T1068Exploitation for Privilege Escalationprivilege escalation85 % - Impacto principal
T1078.001Default Accountsstealth · persistence · privilege escalation · initial access90 %
Acceso local sin interacción (AV:L/UI:N) permite extraer credenciales hardcodeadas en texto plano para obtener acceso con privilegios de clínico (CWE-798: credenciales hardcodeadas).
Inferido por nuestro agente de análisis a partir de la descripción oficial, el vector CVSS y la CWE, y comprobado por un supervisor. Puede contener errores.
🛡️ Mitigaciones ATT&CK que cubren estas técnicas
Tecnologías afectadas (1)
⚠ Inferidas por IA a partir de la descripción — NVD aún no ha analizado esta CVE; no son CPE verificados.
CWE
- CWE-798
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2024-48971",
"cveTags": [],
"metrics": {
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2024-48971",
"role": "CISA Coordinator",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "total"
}
],
"version": "2.0.3",
"timestamp": "2024-11-15T21:05:32.910003Z"
}
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "productsecurity@baxter.com",
"cvssData": {
"scope": "CHANGED",
"version": "3.1",
"baseScore": 9.3,
"attackVector": "LOCAL",
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 6,
"exploitabilityScore": 2.5
}
]
},
"affected": [
{
"source": "productsecurity@baxter.com",
"affectedData": [
{
"vendor": "Baxter",
"product": "Life2000 Ventilation System",
"versions": [
{
"status": "affected",
"version": "06.08.00.00 and prior"
}
],
"defaultStatus": "unaffected"
}
]
},
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"affectedData": [
{
"cpes": [
"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
],
"vendor": "baxter",
"product": "life2000_ventilator_firmware",
"versions": [
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "06.08.00.00"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2024-11-14T22:15:18.327",
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
"source": "productsecurity@baxter.com"
}
],
"vulnStatus": "Deferred",
"weaknesses": [
{
"type": "Secondary",
"source": "productsecurity@baxter.com",
"description": [
{
"lang": "en",
"value": "CWE-798"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges."
},
{
"lang": "es",
"value": "La contraseña del médico y el número de serie están codificados en el respirador en formato de texto simple. Esto podría permitir que un atacante obtenga la contraseña del respirador y la use para obtener acceso no autorizado al dispositivo con privilegios de médico."
}
],
"lastModified": "2026-06-17T07:59:08.533",
"sourceIdentifier": "productsecurity@baxter.com"
}