CVE-2024-48966

Estado: AplazadaCrítica (10)—

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

🎯 Técnicas ATT&CK

Cómo se explota esta vulnerabilidad y qué consigue el atacante, en el lenguaje de MITRE ATT&CK.

Acceso a red (AV:N, PR:N, UI:N) a herramientas sin autenticación permite obtener información diagnóstica y manipular configuración del ventilador. CWE-306 (falta de mecanismo de autenticación) sustenta acceso sin credenciales válidas a funciones administrativas.

Inferido por nuestro agente de análisis a partir de la descripción oficial, el vector CVSS y la CWE, y comprobado por un supervisor. Puede contener errores.

🛡️ Mitigaciones ATT&CK que cubren estas técnicas

Tecnologías afectadas (1)

⚠ Inferidas por IA a partir de la descripción — NVD aún no ha analizado esta CVE; no son CPE verificados.

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2024-48966",
  "cveTags": [],
  "metrics": {
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2024-48966",
          "role": "CISA Coordinator",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "yes"
            },
            {
              "technicalImpact": "total"
            }
          ],
          "version": "2.0.3",
          "timestamp": "2024-11-15T15:33:11.667501Z"
        }
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "productsecurity@baxter.com",
        "cvssData": {
          "scope": "CHANGED",
          "version": "3.1",
          "baseScore": 10,
          "attackVector": "NETWORK",
          "baseSeverity": "CRITICAL",
          "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 6,
        "exploitabilityScore": 3.9
      }
    ]
  },
  "affected": [
    {
      "source": "productsecurity@baxter.com",
      "affectedData": [
        {
          "vendor": "Baxter",
          "product": "Life2000 Ventilation System",
          "versions": [
            {
              "status": "affected",
              "version": "06.08.00.00 and prior"
            }
          ],
          "defaultStatus": "unaffected"
        }
      ]
    },
    {
      "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
      "affectedData": [
        {
          "cpes": [
            "cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
          ],
          "vendor": "baxter",
          "product": "life2000_ventilator_firmware",
          "versions": [
            {
              "status": "affected",
              "version": "0",
              "lessThan": "6.08.00.00",
              "versionType": "custom"
            }
          ],
          "defaultStatus": "unknown"
        }
      ]
    }
  ],
  "published": "2024-11-14T22:15:17.727",
  "references": [
    {
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
      "source": "productsecurity@baxter.com"
    }
  ],
  "vulnStatus": "Deferred",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "productsecurity@baxter.com",
      "description": [
        {
          "lang": "en",
          "value": "CWE-306"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."
    },
    {
      "lang": "es",
      "value": "Las herramientas de software que utiliza el personal de servicio para probar y calibrar el respirador no admiten la autenticación de usuarios. Un atacante con acceso a la PC de servicio donde están instaladas las herramientas podría obtener información de diagnóstico a través de la herramienta de prueba o manipular la configuración del respirador y el software integrado a través de la herramienta de calibración, sin tener que autenticarse en ninguna de las herramientas. Esto podría dar lugar a una divulgación no autorizada de información o tener efectos no deseados en la configuración y el rendimiento del dispositivo."
    }
  ],
  "lastModified": "2026-06-17T07:59:08.223",
  "sourceIdentifier": "productsecurity@baxter.com"
}