« Volver al listado

CVE-2023-25931

Estado: ModificadaMedia (6.8)—

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (2)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2023-25931",
  "cveTags": [],
  "metrics": {
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2023-25931",
          "role": "CISA Coordinator",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "no"
            },
            {
              "technicalImpact": "partial"
            }
          ],
          "version": "2.0.3",
          "timestamp": "2025-03-07T21:34:21.835531Z"
        }
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "security@medtronic.com",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 6.4,
          "attackVector": "PHYSICAL",
          "baseSeverity": "MEDIUM",
          "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "LOW"
        },
        "impactScore": 5.5,
        "exploitabilityScore": 0.9
      },
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 6.8,
          "attackVector": "PHYSICAL",
          "baseSeverity": "MEDIUM",
          "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 0.9
      }
    ]
  },
  "affected": [
    {
      "source": "security@medtronic.com",
      "affectedData": [
        {
          "vendor": "Medtronic",
          "product": "InsterStim Applications",
          "versions": [
            {
              "status": "affected",
              "version": "Micro Clinician ",
              "lessThan": "A51200",
              "versionType": "custom"
            },
            {
              "status": "affected",
              "version": "InterStim X Clinician",
              "lessThan": "A51300",
              "versionType": "custom"
            }
          ],
          "defaultStatus": "unaffected"
        }
      ]
    }
  ],
  "published": "2023-03-01T19:15:26.047",
  "references": [
    {
      "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic-health-interstim-micro.html",
      "tags": [
        "Vendor Advisory"
      ],
      "source": "security@medtronic.com"
    },
    {
      "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic-health-interstim-micro.html",
      "tags": [
        "Vendor Advisory"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "security@medtronic.com",
      "description": [
        {
          "lang": "en",
          "value": "CWE-620"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-287"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app.  Changes still cannot be made outside of the established therapy parameters of the programmer.  For unauthorized access to occur, an individual would need physical access to the Smart Programmer. \n\n"
    }
  ],
  "lastModified": "2026-06-17T05:42:16.960",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:a:medtronic:interstim_x_clinician:a51300:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "81EEC6A7-21C5-46B5-83BD-C3BF94A6D41B"
            },
            {
              "criteria": "cpe:2.3:a:medtronic:micro_clinician:a51200:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "8BDDEB06-CB9C-4CD4-B0E1-59B870AAC7C1"
            }
          ],
          "operator": "OR"
        }
      ]
    }
  ],
  "sourceIdentifier": "security@medtronic.com"
}