CVE-2020-25187
Estado: ModificadaCrítica (9.8)—
Medtronic MyCareLink Smart 25000 is
vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 9.8
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 3.93%
- Percentil entre todas las CVEs puntuadas: 90
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-122
- CWE-787
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2020-25187",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 10,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
"authentication": "NONE",
"integrityImpact": "COMPLETE",
"accessComplexity": "LOW",
"availabilityImpact": "COMPLETE",
"confidentialityImpact": "COMPLETE"
},
"acInsufInfo": false,
"impactScore": 10,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 10,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"cvssData": {
"scope": "CHANGED",
"version": "3.1",
"baseScore": 8.8,
"attackVector": "ADJACENT_NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "REQUIRED",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 6,
"exploitabilityScore": 2.1
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 9.8,
"attackVector": "NETWORK",
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Medtronic",
"product": "Smart Model 25000 Patient Reader",
"versions": [
{
"status": "affected",
"version": "All versions"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2020-12-14T20:15:12.653",
"references": [
{
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-122"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-787"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Medtronic MyCareLink Smart 25000 is \n\n vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device"
},
{
"lang": "es",
"value": "Todas las versiones de Medtronic MyCareLink Smart 25000 son vulnerables cuando un atacante que obtiene la autenticidad ejecuta un comando de depuración, que se envía al lector causando un desbordamiento de la pila en la región stack del MCL Smart Reader. Un desbordamiento de la pila permite al atacante ejecutar remotamente el código en el MCL Smart Reader, lo que podría conllevar al control del dispositivo."
}
],
"lastModified": "2026-06-17T03:06:32.710",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "0E4540F2-921F-4B45-9C30-D1E3F7BE741F"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:medtronic:mycarelink_smart_model_25000:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "06DAC262-42EB-440C-A2B2-3A24A88C05B0"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}