CVE-2020-12040
Estado: ModificadaCrítica (9.8)—
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 9.8
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.95%
- Percentil entre todas las CVEs puntuadas: 60
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-319
- CWE-319
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2020-12040",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 5,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
"authentication": "NONE",
"integrityImpact": "NONE",
"accessComplexity": "LOW",
"availabilityImpact": "NONE",
"confidentialityImpact": "PARTIAL"
},
"acInsufInfo": false,
"impactScore": 2.9,
"baseSeverity": "MEDIUM",
"obtainAllPrivilege": false,
"exploitabilityScore": 10,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV31": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 9.8,
"attackVector": "NETWORK",
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "n/a",
"product": "Baxter Sigma Spectrum Infusion Pumps",
"versions": [
{
"status": "affected",
"version": "Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24"
}
]
}
]
}
],
"published": "2020-06-29T14:15:11.677",
"references": [
{
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-319"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-319"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack."
},
{
"lang": "es",
"value": "Sigma Spectrum Infusion System versiones v's6.x (modelo 35700BAX) y Baxter Spectrum Infusion System versiones 8.x (modelo 35700BAX2), en la capa de aplicación utilizan un canal de comunicación de texto sin cifrar no autenticado para enviar y recibir el estado del sistema y los datos operativos. Esto podría permitir a un atacante que haya eludido las medidas de seguridad de la red visualizar datos confidenciales no privados o llevar a cabo un ataque de tipo man-in-the-middle"
}
],
"lastModified": "2026-06-17T02:51:19.680",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "0FCC49AB-B0AD-44FF-ACEB-3C26EA4F2C3C",
"versionEndIncluding": "6.05",
"versionStartIncluding": "6.0"
},
{
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "72D0F7CB-3D24-4A8D-826D-ACB20ACBEB1C"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "49E25260-EC14-4E98-A86B-CBBE47E26AE5"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}