« Volver al listado

CVE-2018-8842

Estado: ModificadaAlta (8.8)—

Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors. The Philips e-Alert communication channel is not encrypted which could therefore lead to disclosure of personal contact information and application login credentials from within the same subnet.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (1)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2018-8842",
  "cveTags": [],
  "metrics": {
    "cvssMetricV2": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "version": "2.0",
          "baseScore": 3.3,
          "accessVector": "ADJACENT_NETWORK",
          "vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N",
          "authentication": "NONE",
          "integrityImpact": "NONE",
          "accessComplexity": "LOW",
          "availabilityImpact": "NONE",
          "confidentialityImpact": "PARTIAL"
        },
        "acInsufInfo": false,
        "impactScore": 2.9,
        "baseSeverity": "LOW",
        "obtainAllPrivilege": false,
        "exploitabilityScore": 6.5,
        "obtainUserPrivilege": false,
        "obtainOtherPrivilege": false,
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.0",
          "baseScore": 8.8,
          "attackVector": "ADJACENT_NETWORK",
          "baseSeverity": "HIGH",
          "vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 2.8
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "Philips",
          "product": "e-Alert Unit (non-medical device)",
          "versions": [
            {
              "status": "affected",
              "version": "R2.1 and prior"
            }
          ]
        }
      ]
    }
  ],
  "published": "2018-09-26T19:29:00.690",
  "references": [
    {
      "url": "http://www.securityfocus.com/bid/105194",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
      "tags": [
        "Vendor Advisory"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "http://www.securityfocus.com/bid/105194",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    },
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    },
    {
      "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
      "tags": [
        "Vendor Advisory"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-319"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-319"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors. The Philips e-Alert communication channel is not encrypted which could therefore lead to disclosure of personal contact information and application login credentials from within the same subnet."
    },
    {
      "lang": "es",
      "value": "Philips e-Alert Unit (dispositivo no médico), versiones R2.1 y anteriores. El software transmite datos sensibles o críticos para la seguridad en texto claro en un canal de comunicación que puede ser rastreado por actores no autorizados. El canal de comunicación de Philips e-Alert no está cifrado, lo que podría conducir, por lo tanto, a la divulgación de información personal de contacto y de las ccredenciales de inicio de sesión en la aplicación desde dentro de la misma subred."
    }
  ],
  "lastModified": "2026-06-17T02:05:29.570",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:philips:e-alert_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "A20F1AAF-D947-4713-AB4E-12364592DBA3",
              "versionEndIncluding": "r2.1"
            }
          ],
          "operator": "OR"
        }
      ]
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}