« Volver al listado

CVE-2015-3953

Estado: ModificadaCrítica (9.8)—

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (3)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2015-3953",
  "cveTags": [],
  "metrics": {
    "cvssMetricV2": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "version": "2.0",
          "baseScore": 10,
          "accessVector": "NETWORK",
          "vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
          "authentication": "NONE",
          "integrityImpact": "COMPLETE",
          "accessComplexity": "LOW",
          "availabilityImpact": "COMPLETE",
          "confidentialityImpact": "COMPLETE"
        },
        "acInsufInfo": false,
        "impactScore": 10,
        "baseSeverity": "HIGH",
        "obtainAllPrivilege": false,
        "exploitabilityScore": 10,
        "obtainUserPrivilege": false,
        "obtainOtherPrivilege": false,
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.0",
          "baseScore": 9.8,
          "attackVector": "NETWORK",
          "baseSeverity": "CRITICAL",
          "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 3.9
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "Hospira",
          "product": "Plum A+ Infusion System",
          "versions": [
            {
              "status": "affected",
              "version": "<= 13.4"
            }
          ]
        },
        {
          "vendor": "Hospira",
          "product": "Plum A+3 Infusion System",
          "versions": [
            {
              "status": "affected",
              "version": "<= 13.6"
            }
          ]
        },
        {
          "vendor": "Hospira",
          "product": "Symbiq Infusion System",
          "versions": [
            {
              "status": "affected",
              "version": "<= 3.13"
            }
          ]
        }
      ]
    }
  ],
  "published": "2019-03-25T17:29:00.623",
  "references": [
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-259"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-798"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
    },
    {
      "lang": "es",
      "value": "Las cuentas embebidas podrían ser usadas para acceder a Hospira Plum A+ Infusion System versión 13.4 y anteriores, Plum A+3 Infusion System versión 13.6 y anteriores y Symbiq Infusion System, versión 3.13 y anteriores. Hospira recomienda que los usuarios cierren el Puerto 20/FTP y el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado también el Plum 360 Infusion System que no es vulnerable a este problema."
    }
  ],
  "lastModified": "2026-06-17T00:26:32.007",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:plum_a\\+_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2",
              "versionEndIncluding": "13.4"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:plum_a\\+_infusion_system:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "1DB2490B-0318-4770-BF45-CD7527F15D7F"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:plum_a\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "8536E705-89E8-47CB-9567-6AD65FBA0F1B",
              "versionEndIncluding": "13.6"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:plum_a\\+3_infusion_system:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "423AA561-8E38-4378-814B-1008B96F27A6"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "C6DD5141-72AB-4694-8595-B4BED8EC7773",
              "versionEndIncluding": "3.13"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "5C295E1A-BF60-476D-B972-5C5C28D7633B"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}