« Volver al listado

CVE-2014-7233

Estado: ModificadaAlta (10)—

GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740_122_Setup.exe, (3) hrml for "Setup and Activation" using DSASetup, and (4) an empty string for Shutter Configuration, which has unspecified impact and attack vectors. NOTE: since these passwords appear to be used to access functionality during installation, this issue might not cross privilege boundaries and might not be a vulnerability.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (1)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2014-7233",
  "cveTags": [],
  "metrics": {
    "cvssMetricV2": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "version": "2.0",
          "baseScore": 10,
          "accessVector": "NETWORK",
          "vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
          "authentication": "NONE",
          "integrityImpact": "COMPLETE",
          "accessComplexity": "LOW",
          "availabilityImpact": "COMPLETE",
          "confidentialityImpact": "COMPLETE"
        },
        "acInsufInfo": true,
        "impactScore": 10,
        "baseSeverity": "HIGH",
        "obtainAllPrivilege": false,
        "exploitabilityScore": 10,
        "obtainUserPrivilege": false,
        "obtainOtherPrivilege": false,
        "userInteractionRequired": false
      }
    ]
  },
  "affected": [
    {
      "source": "cve@mitre.org",
      "affectedData": [
        {
          "vendor": "n/a",
          "product": "n/a",
          "versions": [
            {
              "status": "affected",
              "version": "n/a"
            }
          ]
        }
      ]
    }
  ],
  "published": "2015-08-04T14:59:25.720",
  "references": [
    {
      "url": "http://apps.gehealthcare.com/servlet/ClientServlet/5458232-1EN+r4.pdf?REQ=RAA&DIRECTION=5458232-1EN&FILENAME=5458232-1EN%2Br4.pdf&FILEREV=4&DOCREV_ORG=4",
      "source": "cve@mitre.org"
    },
    {
      "url": "http://www.forbes.com/sites/thomasbrewster/2015/07/10/vulnerable-breasts/",
      "source": "cve@mitre.org"
    },
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-037-02",
      "source": "cve@mitre.org"
    },
    {
      "url": "https://twitter.com/digitalbond/status/619250429751222277",
      "source": "cve@mitre.org"
    },
    {
      "url": "http://apps.gehealthcare.com/servlet/ClientServlet/5458232-1EN+r4.pdf?REQ=RAA&DIRECTION=5458232-1EN&FILENAME=5458232-1EN%2Br4.pdf&FILEREV=4&DOCREV_ORG=4",
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    },
    {
      "url": "http://www.forbes.com/sites/thomasbrewster/2015/07/10/vulnerable-breasts/",
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    },
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-037-02",
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    },
    {
      "url": "https://twitter.com/digitalbond/status/619250429751222277",
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-255"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740_122_Setup.exe, (3) hrml for \"Setup and Activation\" using DSASetup, and (4) an empty string for Shutter Configuration, which has unspecified impact and attack vectors.  NOTE: since these passwords appear to be used to access functionality during installation, this issue might not cross privilege boundaries and might not be a vulnerability."
    },
    {
      "lang": "es",
      "value": "Vulnerabilidad en GE Healthcare Precisión THUNIS-800+, tiene una contraseña predeterminada de (1) 1973 para el menú System Utilities predeterminado de fábrica, (2) TH8740 para la instalación usando TH8740_122_Setup.exe, (3) hrml para 'Setup and Activation' usando DSASetup y (4) una cadena vacía para Shutter Configuration, lo cual tiene un impacto y vectores de ataque no especificados. NOTA: ya que estas contraseñas parecen ser utilizadas para tener acceso a la funcionalidad durante la instalación, este problema podría no cruzar los límites de los privilegios y podría no ser una vulnerabilidad."
    }
  ],
  "lastModified": "2026-06-17T00:14:36.243",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:gehealthcare:precision_thunis-800\\+:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "DEF95E63-0F28-4644-B98A-1AB28399923F"
            }
          ],
          "operator": "OR"
        }
      ]
    }
  ],
  "sourceIdentifier": "cve@mitre.org"
}