Baxter Sigma Spectrum 35700bax Firmware: vulnerabilities and CVEs
Baxter Sigma Spectrum 35700bax Firmware has 4 published vulnerabilities, 0 of them in the last 12 months. 0 are rated critical and 0 are listed by CISA as actively exploited.
CVEs4
Last 12 months0
Critical0
Actively exploited0
All vulnerabilities in the catalogue →⭐ Follow this technology
Latest vulnerabilities
| CVE | Severity | EPSS | Active exploitation | Published | Description |
|---|---|---|---|---|---|
| CVE-2022-26394 | Medium (5.4) | 0.35% | — | Sep 9, 2022 | The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. This may allow an attacker to perform a man in the middle attack that modifies parameters making the network connection fail. |
| CVE-2022-26393 | High (8.1) | 0.67% | — | Sep 9, 2022 | The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information or cause a Denial of Service (DoS) on the… |
| CVE-2022-26392 | Medium (6.5) | 0.66% | — | Sep 9, 2022 | The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. An attacker could use this to read… |
| CVE-2022-26390 | Medium (4.2) | 0.47% | — | Sep 9, 2022 | The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that… |
Other products by Baxter
Sigma Spectrum Infusion System Firmware · 10Em2400 Firmware · 6Em1200 Firmware · 6Baxter Spectrum IQ 35700bax3 Firmware · 4Sigma Spectrum 35700bax2 Firmware · 4Spectrum Wireless Battery Module Firmware · 4Prismax Firmware · 3Prismaflex Firmware · 3Connex Health Portal · 2Welch Allyn Xscribe Cardiac Stress Testing System Firmware · 1Welch Allyn Configuration Tool · 1Welch Allyn Connex Cardio · 1