CVE-2025-4393
Estado: AplazadaMedia (6.5)—
Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges.
This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H
- Puntuación base: 6.5
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.17%
- Percentil entre todas las CVEs puntuadas: 6
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
⚠ Inferidas por IA a partir de la descripción — NVD aún no ha analizado esta CVE; no son CPE verificados.
CWE
- CWE-502
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2025-4393",
"cveTags": [],
"metrics": {
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2025-4393",
"role": "CISA Coordinator",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "total"
}
],
"version": "2.0.3",
"timestamp": "2025-07-24T14:33:13.440835Z"
}
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "security@medtronic.com",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 6.5,
"attackVector": "LOCAL",
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "HIGH",
"availabilityImpact": "HIGH",
"privilegesRequired": "LOW",
"confidentialityImpact": "LOW"
},
"impactScore": 5.5,
"exploitabilityScore": 1
}
]
},
"affected": [
{
"source": "security@medtronic.com",
"affectedData": [
{
"vendor": "Medtronic",
"product": "MyCareLink Patient Monitor 24950",
"versions": [
{
"status": "affected",
"version": "0",
"lessThan": "June 25, 2025",
"versionType": "custom"
}
],
"defaultStatus": "unaffected"
},
{
"vendor": "Medtronic",
"product": "MyCareLink Patient Monitor 24952",
"versions": [
{
"status": "affected",
"version": "0",
"lessThan": "June 25, 2025",
"versionType": "custom"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2025-07-24T07:15:53.343",
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01",
"source": "security@medtronic.com"
},
{
"url": "https://www.medtronic.com/en-us/e/product-security/security-bulletins/mycarelink-patient-monitor-vulnerabilities.html",
"source": "security@medtronic.com"
}
],
"vulnStatus": "Deferred",
"weaknesses": [
{
"type": "Secondary",
"source": "security@medtronic.com",
"description": [
{
"lang": "en",
"value": "CWE-502"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges. \n\nThis issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025"
},
{
"lang": "es",
"value": "Medtronic MyCareLink Patient Monitor cuenta con un servicio interno que deserializa datos, lo que permite a un atacante local interactuar con el servicio mediante la manipulado de una carga binaria para bloquearlo o elevar privilegios. Este problema afecta a los modelos 24950 y 24952 de Medtronic MyCareLink Patient Monitor antes del 25 de junio de 2025."
}
],
"lastModified": "2026-06-17T09:33:10.723",
"sourceIdentifier": "security@medtronic.com"
}