CVE-2025-14097

Estado: AplazadaAlta (7.2)—

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.

Leer descripción completaMostrar menos

Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status:

Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

🎯 Técnicas ATT&CK

Cómo se explota esta vulnerabilidad y qué consigue el atacante, en el lenguaje de MITRE ATT&CK.

Vector CVSS con PR:H y AV:N indica acceso remoto privilegiado a servicio. CWE-287 (autenticación débil) + 'remote support feature' sugiere T1210. Impactos: acceso no autorizado (T1078) y ejecución de código remota mencionada explícitamente.

Inferido por nuestro agente de análisis a partir de la descripción oficial, el vector CVSS y la CWE, y comprobado por un supervisor. Puede contener errores.

🛡️ Mitigaciones ATT&CK que cubren estas técnicas

Tecnologías afectadas (1)

⚠ Inferidas por IA a partir de la descripción — NVD aún no ha analizado esta CVE; no son CPE verificados.

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2025-14097",
  "cveTags": [],
  "metrics": {
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2025-14097",
          "role": "CISA Coordinator",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "no"
            },
            {
              "technicalImpact": "total"
            }
          ],
          "version": "2.0.3",
          "timestamp": "2025-12-17T14:40:30.048229Z"
        }
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 7.2,
          "attackVector": "NETWORK",
          "baseSeverity": "HIGH",
          "vectorString": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "HIGH",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 1.2
      }
    ]
  },
  "affected": [
    {
      "source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
      "affectedData": [
        {
          "vendor": "Radiometer Medical Aps",
          "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
          "versions": [
            {
              "status": "affected",
              "version": "Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems",
              "versionType": "ABL90 DMS (Data Management system)"
            },
            {
              "status": "unaffected",
              "version": "Application software versions >= 3.5MR11 with Windows 10 operating system",
              "versionType": "ABL90 DMS (Data Management system)"
            }
          ],
          "defaultStatus": "unaffected"
        },
        {
          "vendor": "Radiometer Medical Aps",
          "product": "ABL800 BASIC and ABL800 FLEX Analyzers",
          "versions": [
            {
              "status": "affected",
              "version": "Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems",
              "versionType": "ABL800 DMS(Data Management System)"
            },
            {
              "status": "unaffected",
              "version": "Application software versions >= 6.20MR2 with Windows 10 operating system",
              "versionType": "ABL800 DMS(Data Management System)"
            }
          ],
          "defaultStatus": "unaffected"
        },
        {
          "vendor": "Radiometer Medical Aps",
          "product": "AQT90 FLEX Analyzers",
          "versions": [
            {
              "status": "affected",
              "version": "Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems",
              "versionType": "AQT90 DMS (Data Management System)"
            },
            {
              "status": "unaffected",
              "version": "Application software versions >= 8.13 MR2 with Windows 10 operating system",
              "versionType": "AQT90 DMS (Data Management System)"
            }
          ],
          "defaultStatus": "unaffected"
        }
      ]
    }
  ],
  "published": "2025-12-17T13:15:58.050",
  "references": [
    {
      "url": "https://www.radiometer.com/myradiometer",
      "source": "46b595e9-1acc-41cb-9398-adaf98d37a9b"
    }
  ],
  "vulnStatus": "Deferred",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
      "description": [
        {
          "lang": "en",
          "value": "CWE-287"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means.  The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\nRequired Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:\nCustomers should ensure the following:\n•  The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     \nExploit Status:\n\n\nResearchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication."
    }
  ],
  "lastModified": "2026-06-17T08:35:19.363",
  "sourceIdentifier": "46b595e9-1acc-41cb-9398-adaf98d37a9b"
}