CVE-2025-14097
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.
Leer descripción completaMostrar menos
Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status:
Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 7.2
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.46%
- Percentil entre todas las CVEs puntuadas: 38
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
🎯 Técnicas ATT&CK
Cómo se explota esta vulnerabilidad y qué consigue el atacante, en el lenguaje de MITRE ATT&CK.
- Explotación
T1210Exploitation of Remote Serviceslateral movement85 % - Impacto principal
T1078Valid Accountsstealth · persistence · privilege escalation · initial access75 % - Impacto secundario
T1059Command and Scripting Interpreterexecution70 %
Vector CVSS con PR:H y AV:N indica acceso remoto privilegiado a servicio. CWE-287 (autenticación débil) + 'remote support feature' sugiere T1210. Impactos: acceso no autorizado (T1078) y ejecución de código remota mencionada explícitamente.
Inferido por nuestro agente de análisis a partir de la descripción oficial, el vector CVSS y la CWE, y comprobado por un supervisor. Puede contener errores.
🛡️ Mitigaciones ATT&CK que cubren estas técnicas
Tecnologías afectadas (1)
⚠ Inferidas por IA a partir de la descripción — NVD aún no ha analizado esta CVE; no son CPE verificados.
CWE
- CWE-287
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2025-14097",
"cveTags": [],
"metrics": {
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2025-14097",
"role": "CISA Coordinator",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "total"
}
],
"version": "2.0.3",
"timestamp": "2025-12-17T14:40:30.048229Z"
}
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 7.2,
"attackVector": "NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "HIGH",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 1.2
}
]
},
"affected": [
{
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"affectedData": [
{
"vendor": "Radiometer Medical Aps",
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"versions": [
{
"status": "affected",
"version": "Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems",
"versionType": "ABL90 DMS (Data Management system)"
},
{
"status": "unaffected",
"version": "Application software versions >= 3.5MR11 with Windows 10 operating system",
"versionType": "ABL90 DMS (Data Management system)"
}
],
"defaultStatus": "unaffected"
},
{
"vendor": "Radiometer Medical Aps",
"product": "ABL800 BASIC and ABL800 FLEX Analyzers",
"versions": [
{
"status": "affected",
"version": "Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems",
"versionType": "ABL800 DMS(Data Management System)"
},
{
"status": "unaffected",
"version": "Application software versions >= 6.20MR2 with Windows 10 operating system",
"versionType": "ABL800 DMS(Data Management System)"
}
],
"defaultStatus": "unaffected"
},
{
"vendor": "Radiometer Medical Aps",
"product": "AQT90 FLEX Analyzers",
"versions": [
{
"status": "affected",
"version": "Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems",
"versionType": "AQT90 DMS (Data Management System)"
},
{
"status": "unaffected",
"version": "Application software versions >= 8.13 MR2 with Windows 10 operating system",
"versionType": "AQT90 DMS (Data Management System)"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2025-12-17T13:15:58.050",
"references": [
{
"url": "https://www.radiometer.com/myradiometer",
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b"
}
],
"vulnStatus": "Deferred",
"weaknesses": [
{
"type": "Secondary",
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"description": [
{
"lang": "en",
"value": "CWE-287"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\nRequired Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution:\nCustomers should ensure the following:\n• The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution. \nExploit Status:\n\n\nResearchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication."
}
],
"lastModified": "2026-06-17T08:35:19.363",
"sourceIdentifier": "46b595e9-1acc-41cb-9398-adaf98d37a9b"
}