CVE-2023-0888
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 7.2
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.60%
- Percentil entre todas las CVEs puntuadas: 47
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-95
- CWE-94
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2023-0888",
"cveTags": [],
"metrics": {
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2023-0888",
"role": "CISA Coordinator",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "partial"
}
],
"version": "2.0.3",
"timestamp": "2025-02-27T18:17:34.759625Z"
}
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "productsecurity@bbraun.com",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 4.9,
"attackVector": "NETWORK",
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "NONE",
"privilegesRequired": "HIGH",
"confidentialityImpact": "NONE"
},
"impactScore": 3.6,
"exploitabilityScore": 1.2
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 7.2,
"attackVector": "NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "HIGH",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 1.2
}
]
},
"affected": [
{
"source": "productsecurity@bbraun.com",
"affectedData": [
{
"vendor": "B. Braun ",
"product": "Space Battery Pack SP with Wi-Fi",
"versions": [
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "053L000092"
},
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "054U000092"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2023-03-13T09:15:10.013",
"references": [
{
"url": "https://www.bbraun.com/productsecurity",
"tags": [
"Product"
],
"source": "productsecurity@bbraun.com"
},
{
"url": "https://www.bbraunusa.com/productsecurity",
"tags": [
"Product"
],
"source": "productsecurity@bbraun.com"
},
{
"url": "https://www.bbraun.com/productsecurity",
"tags": [
"Product"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://www.bbraunusa.com/productsecurity",
"tags": [
"Product"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "productsecurity@bbraun.com",
"description": [
{
"lang": "en",
"value": "CWE-95"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-94"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks\n\n"
}
],
"lastModified": "2026-06-17T05:26:33.543",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:global:*:*:*",
"vulnerable": true,
"matchCriteriaId": "8B1E3BFE-FA53-455B-8123-5832752563D6",
"versionEndIncluding": "053l000092"
},
{
"criteria": "cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:us:*:*:*",
"vulnerable": true,
"matchCriteriaId": "52E526B0-39B2-42DB-B00C-5FC80351A082",
"versionEndIncluding": "054u000092"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:bbraun:battery-pack_sp_with_wifi:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "62C8A903-575E-42A3-8980-ABE788309AD4"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "productsecurity@bbraun.com"
}