CVE-2022-38100
Estado: ModificadaAlta (7.5)—
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
- Puntuación base: 7.5
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.98%
- Percentil entre todas las CVEs puntuadas: 61
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-400
- NVD-CWE-Other
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2022-38100",
"cveTags": [],
"metrics": {
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2022-38100",
"role": "CISA Coordinator",
"options": [
{
"exploitation": "none"
},
{
"automatable": "yes"
},
{
"technicalImpact": "partial"
}
],
"version": "2.0.3",
"timestamp": "2025-04-16T15:54:20.725304Z"
}
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 7.5,
"attackVector": "NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"integrityImpact": "NONE",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "NONE"
},
"impactScore": 3.6,
"exploitabilityScore": 3.9
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 7.5,
"attackVector": "NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"integrityImpact": "NONE",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "NONE"
},
"impactScore": 3.6,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Contec Health",
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
]
}
],
"published": "2022-09-13T15:15:08.843",
"references": [
{
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-400"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "NVD-CWE-Other"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
},
{
"lang": "es",
"value": "El dispositivo CMS800 falla mientras intenta analizar los datos de red malformados enviados por un actor de la amenaza. Un actor de la amenaza con acceso a la red puede emitir de forma remota una petición UDP con un formato especial que causará el bloqueo de todo el dispositivo y requerirá un reinicio físico. Puede enviarse una petición de difusión UDP que cause un ataque de denegación de servicio masivo en todos los dispositivos CME8000 conectados a la misma red"
}
],
"lastModified": "2026-06-17T04:56:06.327",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}