« Volver al listado

CVE-2022-38069

Estado: ModificadaMedia (6.1)—

Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (1)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2022-38069",
  "cveTags": [],
  "metrics": {
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2022-38069",
          "role": "CISA Coordinator",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "no"
            },
            {
              "technicalImpact": "partial"
            }
          ],
          "version": "2.0.3",
          "timestamp": "2025-04-16T15:50:05.918933Z"
        }
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "ics-cert@hq.dhs.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 4.3,
          "attackVector": "PHYSICAL",
          "baseSeverity": "MEDIUM",
          "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
          "integrityImpact": "LOW",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "LOW",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "LOW"
        },
        "impactScore": 3.4,
        "exploitabilityScore": 0.9
      },
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 6.1,
          "attackVector": "PHYSICAL",
          "baseSeverity": "MEDIUM",
          "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "NONE",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.2,
        "exploitabilityScore": 0.9
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "Contec Health",
          "product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
          "versions": [
            {
              "status": "affected",
              "version": "All"
            }
          ]
        }
      ]
    }
  ],
  "published": "2022-09-13T15:15:08.777",
  "references": [
    {
      "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-798"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-798"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
    },
    {
      "lang": "es",
      "value": "Se presentan múltiples credenciales predeterminadas globales en todos los dispositivos CMS8000 que, una vez expuestas, permiten a un actor de amenazas con acceso físico momentáneo conseguir acceso privilegiado a cualquier dispositivo. El acceso con credenciales privilegiadas permite una extracción de información confidencial del paciente o la modificación de los parámetros del dispositivo"
    }
  ],
  "lastModified": "2026-06-17T04:56:02.110",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}