CVE-2021-37164
Estado: ModificadaCrítica (9.8)—
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 9.8
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 3.40%
- Percentil entre todas las CVEs puntuadas: 88
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-787
Referencias
- https://www.armis.com/PwnedPiper
- https://www.swisslog-healthcare.com
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20
- https://www.armis.com/PwnedPiper
- https://www.swisslog-healthcare.com
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20
JSON original (NVD)
Mostrar
{
"id": "CVE-2021-37164",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 7.5,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:L/Au:N/C:P/I:P/A:P",
"authentication": "NONE",
"integrityImpact": "PARTIAL",
"accessComplexity": "LOW",
"availabilityImpact": "PARTIAL",
"confidentialityImpact": "PARTIAL"
},
"acInsufInfo": false,
"impactScore": 6.4,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 10,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV31": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 9.8,
"attackVector": "NETWORK",
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "cve@mitre.org",
"affectedData": [
{
"vendor": "n/a",
"product": "n/a",
"versions": [
{
"status": "affected",
"version": "n/a"
}
]
}
]
}
],
"published": "2021-08-02T13:15:07.830",
"references": [
{
"url": "https://www.armis.com/PwnedPiper",
"tags": [
"Broken Link"
],
"source": "cve@mitre.org"
},
{
"url": "https://www.swisslog-healthcare.com",
"tags": [
"Product"
],
"source": "cve@mitre.org"
},
{
"url": "https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741",
"tags": [
"Vendor Advisory"
],
"source": "cve@mitre.org"
},
{
"url": "https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20",
"source": "cve@mitre.org"
},
{
"url": "https://www.armis.com/PwnedPiper",
"tags": [
"Broken Link"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://www.swisslog-healthcare.com",
"tags": [
"Product"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741",
"tags": [
"Vendor Advisory"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20",
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-787"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow."
},
{
"lang": "es",
"value": "Se ha detectado un problema de desbordamiento de búfer en el panel de control HMI3 en el Panel Swisslog Healthcare Nexus que funciona con versiones de Nexus Software anteriores a 7.2.5.7. En la función tcpTxThread, los datos recibidos se copian en un búfer de pila. Puede ocurrir una condición off-by-3, resultando en un desbordamiento del búfer en la región stack de la memoria"
}
],
"lastModified": "2026-06-17T04:00:10.627",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:swisslog-healthcare:hmi-3_control_panel_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "B9286DC1-0111-49C4-9BD4-1601AAD025A9",
"versionEndExcluding": "7.2.5.7"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:swisslog-healthcare:hmi-3_control_panel:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "92F48887-6E68-4A96-87F6-C8DB7773C4A2"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "cve@mitre.org"
}