CVE-2020-27252
Estado: ModificadaAlta (8.1)—
Medtronic MyCareLink Smart 25000 is
vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 8.1
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 3.84%
- Percentil entre todas las CVEs puntuadas: 90
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-367
- CWE-367
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2020-27252",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 9.3,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:M/Au:N/C:C/I:C/A:C",
"authentication": "NONE",
"integrityImpact": "COMPLETE",
"accessComplexity": "MEDIUM",
"availabilityImpact": "COMPLETE",
"confidentialityImpact": "COMPLETE"
},
"acInsufInfo": false,
"impactScore": 10,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 8.6,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"cvssData": {
"scope": "CHANGED",
"version": "3.1",
"baseScore": 8.8,
"attackVector": "ADJACENT_NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "REQUIRED",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 6,
"exploitabilityScore": 2.1
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 8.1,
"attackVector": "NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "HIGH",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 2.2
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Medtronic",
"product": "Smart Model 25000 Patient Reader",
"versions": [
{
"status": "affected",
"version": "All versions"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2020-12-14T20:15:12.717",
"references": [
{
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-367"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-367"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Medtronic MyCareLink Smart 25000 is \n\nvulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device."
},
{
"lang": "es",
"value": "Todas las versiones de Medtronic MyCareLink Smart 25000 son vulnerables a una condición de carrera en el sistema de actualización del software MCL Smart Patient Reader, que permite que el firmware sin firmar sea cargado y ejecutado en el Patient Reader. Si es explotado, un atacante podría ejecutar remotamente el código en el dispositivo MCL Smart Patient Reader, conllevando a un control del dispositivo"
}
],
"lastModified": "2026-06-17T03:09:03.333",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "0E4540F2-921F-4B45-9C30-D1E3F7BE741F"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:medtronic:mycarelink_smart_model_25000:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "06DAC262-42EB-440C-A2B2-3A24A88C05B0"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}