« Volver al listado

CVE-2020-12037

Estado: ModificadaAlta (7.5)—

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (2)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2020-12037",
  "cveTags": [],
  "metrics": {
    "cvssMetricV2": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "version": "2.0",
          "baseScore": 5,
          "accessVector": "NETWORK",
          "vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
          "authentication": "NONE",
          "integrityImpact": "NONE",
          "accessComplexity": "LOW",
          "availabilityImpact": "NONE",
          "confidentialityImpact": "PARTIAL"
        },
        "acInsufInfo": false,
        "impactScore": 2.9,
        "baseSeverity": "MEDIUM",
        "obtainAllPrivilege": false,
        "exploitabilityScore": 10,
        "obtainUserPrivilege": false,
        "obtainOtherPrivilege": false,
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 7.5,
          "attackVector": "NETWORK",
          "baseSeverity": "HIGH",
          "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
          "integrityImpact": "NONE",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "NONE",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 3.6,
        "exploitabilityScore": 3.9
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "n/a",
          "product": "Baxter PrismaFlex and PrisMax",
          "versions": [
            {
              "status": "affected",
              "version": "PrismaFlex all versions, PrisMax all versions prior to 3.x"
            }
          ]
        }
      ]
    }
  ],
  "published": "2020-06-29T14:15:11.537",
  "references": [
    {
      "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
      "tags": [
        "Not Applicable"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02",
      "tags": [
        "Third Party Advisory"
      ],
      "source": "nvd@nist.gov"
    },
    {
      "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
      "tags": [
        "Not Applicable"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-259"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-319"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device."
    },
    {
      "lang": "es",
      "value": "Baxter PrismaFlex todas las versiones, PrisMax todas las versiones anteriores a 3.x, los dispositivos afectados no implementan cifrado de datos en tránsito (por ejemplo, TLS/SSL) cuando se configuran para enviar datos de tratamiento a un PDMS (Patient Data Management System) o un sistema EMR (Electronic Medical Record). Un atacante podría observar datos confidenciales enviados desde el dispositivo"
    }
  ],
  "lastModified": "2026-06-17T02:51:18.987",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:baxter:prismaflex_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "B8874804-02E2-4E74-94C6-F103DCDC854E"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:baxter:prismaflex:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "8E5DB2B2-5D9E-494B-A860-FE2F4BC4C767"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:baxter:prismax_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "8F197DAD-63A3-4598-BBFC-0014707B3A01",
              "versionEndExcluding": "3.0"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:baxter:prismax:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "A5EB1A30-A514-437D-B976-451CF003955D"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "evaluatorComment": "At the time of analysis, information provided did not align with the data (including CVE number) in the advisory.",
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}