CVE-2020-12036
Estado: ModificadaAlta (7.5)—
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- Puntuación base: 7.5
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.50%
- Percentil entre todas las CVEs puntuadas: 40
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (2)
CWE
- CWE-319
- CWE-319
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2020-12036",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 5,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
"authentication": "NONE",
"integrityImpact": "NONE",
"accessComplexity": "LOW",
"availabilityImpact": "NONE",
"confidentialityImpact": "PARTIAL"
},
"acInsufInfo": false,
"impactScore": 2.9,
"baseSeverity": "MEDIUM",
"obtainAllPrivilege": false,
"exploitabilityScore": 10,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV31": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 7.5,
"attackVector": "NETWORK",
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
"integrityImpact": "NONE",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "NONE",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 3.6,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "n/a",
"product": "Baxter PrismaFlex and PrisMax",
"versions": [
{
"status": "affected",
"version": "PrismaFlex all versions, PrisMax all versions prior to 3.x"
}
]
}
]
}
],
"published": "2020-06-29T14:15:11.460",
"references": [
{
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
"tags": [
"Not Applicable"
],
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
"tags": [
"Not Applicable"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-319"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-319"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device."
},
{
"lang": "es",
"value": "Baxter PrismaFlex todas las versiones, PrisMax todas las versiones anteriores a 3.x, los dispositivos afectados no implementan cifrado de datos en tránsito (por ejemplo, TLS/SSL) cuando se configuran para enviar datos de tratamiento a un PDMS (Patient Data Management System) o un sistema EMR (Electronic Medical Record). Un atacante podría observar datos confidenciales enviados desde el dispositivo"
}
],
"lastModified": "2026-06-17T02:51:18.673",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:baxter:prismaflex_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "B8874804-02E2-4E74-94C6-F103DCDC854E"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:baxter:prismaflex:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "8E5DB2B2-5D9E-494B-A860-FE2F4BC4C767"
}
],
"operator": "OR"
}
],
"operator": "AND"
},
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:baxter:prismax_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "8F197DAD-63A3-4598-BBFC-0014707B3A01",
"versionEndExcluding": "3.0"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:baxter:prismax:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "A5EB1A30-A514-437D-B976-451CF003955D"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}