« Volver al listado

CVE-2019-13535

Estado: ModificadaMedia (4.6)—

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (2)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2019-13535",
  "cveTags": [],
  "metrics": {
    "cvssMetricV2": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "version": "2.0",
          "baseScore": 2.1,
          "accessVector": "LOCAL",
          "vectorString": "AV:L/AC:L/Au:N/C:P/I:N/A:N",
          "authentication": "NONE",
          "integrityImpact": "NONE",
          "accessComplexity": "LOW",
          "availabilityImpact": "NONE",
          "confidentialityImpact": "PARTIAL"
        },
        "acInsufInfo": false,
        "impactScore": 2.9,
        "baseSeverity": "LOW",
        "obtainAllPrivilege": false,
        "exploitabilityScore": 3.9,
        "obtainUserPrivilege": false,
        "obtainOtherPrivilege": false,
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV31": [
      {
        "type": "Secondary",
        "source": "ics-cert@hq.dhs.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 4.6,
          "attackVector": "PHYSICAL",
          "baseSeverity": "MEDIUM",
          "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
          "integrityImpact": "NONE",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "NONE",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 3.6,
        "exploitabilityScore": 0.9
      },
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.1",
          "baseScore": 4.6,
          "attackVector": "PHYSICAL",
          "baseSeverity": "MEDIUM",
          "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
          "integrityImpact": "NONE",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "NONE",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 3.6,
        "exploitabilityScore": 0.9
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "Medtronic",
          "product": "Valleylab FT10 Energy Platform (VLFT10GEN)",
          "versions": [
            {
              "status": "affected",
              "version": "0",
              "versionType": "custom",
              "lessThanOrEqual": "2.1.0"
            },
            {
              "status": "affected",
              "version": "0",
              "versionType": "custom",
              "lessThanOrEqual": "2.0.3"
            }
          ],
          "defaultStatus": "unaffected"
        },
        {
          "vendor": "Medtronic",
          "product": "Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)",
          "versions": [
            {
              "status": "affected",
              "version": "0",
              "versionType": "custom",
              "lessThanOrEqual": "1.20.2"
            }
          ],
          "defaultStatus": "unaffected"
        }
      ]
    }
  ],
  "published": "2019-11-08T20:15:10.633",
  "references": [
    {
      "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html",
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01",
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-693"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-732"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data."
    },
    {
      "lang": "es",
      "value": "En Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) versión 2.1.0 y versiones inferiores y versión 2.0.3 y versiones inferiores, y Valleylab LS10 Energy Platform (VLLS10GEN —no disponible en los Estados Unidos) versión 1.20.2 y versiones inferiores, el mecanismo de seguridad RFID sí no aplica protección de lectura, lo que permite el acceso de lectura total de los datos del mecanismo de seguridad RFID"
    }
  ],
  "lastModified": "2026-06-17T02:16:55.827",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.0.3:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "89DAFB01-536A-4966-8EAC-5D6CED32CC63"
            },
            {
              "criteria": "cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.1.0:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "31DF6D35-968D-4E8D-A8F5-578B4B9C2BE9"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronic:valleylab_ft10_energy_platform:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "164230CB-E2BF-447F-8537-C9401FA0CC09"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronic:valleylab_ls10_energy_platform_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "EEAE74E9-49AC-4B77-8831-11EA09B204CB",
              "versionEndIncluding": "1.20.2"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronic:valleylab_ls10_energy_platform:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "82294F1E-0BD0-4F55-A732-09B68AA37624"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}