CVE-2019-13531
Estado: ModificadaMedia (4.6)—
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
- Puntuación base: 4.6
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.39%
- Percentil entre todas las CVEs puntuadas: 31
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (2)
CWE
- CWE-287
- NVD-CWE-noinfo
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2019-13531",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 2.1,
"accessVector": "LOCAL",
"vectorString": "AV:L/AC:L/Au:N/C:N/I:P/A:N",
"authentication": "NONE",
"integrityImpact": "PARTIAL",
"accessComplexity": "LOW",
"availabilityImpact": "NONE",
"confidentialityImpact": "NONE"
},
"acInsufInfo": false,
"impactScore": 2.9,
"baseSeverity": "LOW",
"obtainAllPrivilege": false,
"exploitabilityScore": 3.9,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 4.8,
"attackVector": "PHYSICAL",
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "HIGH",
"availabilityImpact": "LOW",
"privilegesRequired": "NONE",
"confidentialityImpact": "NONE"
},
"impactScore": 4.2,
"exploitabilityScore": 0.5
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 4.6,
"attackVector": "PHYSICAL",
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "NONE",
"privilegesRequired": "NONE",
"confidentialityImpact": "NONE"
},
"impactScore": 3.6,
"exploitabilityScore": 0.9
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Medtronic",
"product": "Valleylab FT10 Energy Platform (VLFT10GEN)",
"versions": [
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "2.1.0"
},
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "2.0.3"
}
],
"defaultStatus": "unaffected"
},
{
"vendor": "Medtronic",
"product": "Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)",
"versions": [
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "1.20.2"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2019-11-08T20:15:10.570",
"references": [
{
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-287"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "NVD-CWE-noinfo"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
},
{
"lang": "es",
"value": "En Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) versión 2.1.0 y versiones anteriores y versión 2.0.3 y versiones atneriores, y Valleylab LS10 Energy Platform (VLLS10GEN —no disponible en los Estados Unidos) versión 1.20.2 y versiones anteriores, se utilizó el mecanismo de seguridad RFID para la autenticación entre la plataforma de energía FT10 / LS10 y los instrumentos se pueden omitir, lo que permite que los instrumentos no auténticos se conecten al generador."
}
],
"lastModified": "2026-06-17T02:16:55.257",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.0.3:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "89DAFB01-536A-4966-8EAC-5D6CED32CC63"
},
{
"criteria": "cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.1.0:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "31DF6D35-968D-4E8D-A8F5-578B4B9C2BE9"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:medtronic:valleylab_ft10_energy_platform:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "164230CB-E2BF-447F-8537-C9401FA0CC09"
}
],
"operator": "OR"
}
],
"operator": "AND"
},
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:medtronic:valleylab_ls10_energy_platform_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "EEAE74E9-49AC-4B77-8831-11EA09B204CB",
"versionEndIncluding": "1.20.2"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:medtronic:valleylab_ls10_energy_platform:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "82294F1E-0BD0-4F55-A732-09B68AA37624"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}