CVE-2018-5448
Estado: ModificadaMedia (4.8)—
Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system.
CVSS
- Versión: 3.1
- Vector: CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:H/I:N/A:N
- Puntuación base: 4.8
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 0.74%
- Percentil entre todas las CVEs puntuadas: 53
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-23
- CWE-22
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2018-5448",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 2.7,
"accessVector": "ADJACENT_NETWORK",
"vectorString": "AV:A/AC:L/Au:S/C:P/I:N/A:N",
"authentication": "SINGLE",
"integrityImpact": "NONE",
"accessComplexity": "LOW",
"availabilityImpact": "NONE",
"confidentialityImpact": "PARTIAL"
},
"acInsufInfo": false,
"impactScore": 2.9,
"baseSeverity": "LOW",
"obtainAllPrivilege": false,
"exploitabilityScore": 5.1,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV30": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.0",
"baseScore": 5.7,
"attackVector": "ADJACENT_NETWORK",
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:N",
"integrityImpact": "NONE",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "NONE",
"privilegesRequired": "LOW",
"confidentialityImpact": "HIGH"
},
"impactScore": 3.6,
"exploitabilityScore": 2.1
}
],
"cvssMetricV31": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.1",
"baseScore": 4.8,
"attackVector": "ADJACENT_NETWORK",
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:H/I:N/A:N",
"integrityImpact": "NONE",
"userInteraction": "NONE",
"attackComplexity": "HIGH",
"availabilityImpact": "NONE",
"privilegesRequired": "LOW",
"confidentialityImpact": "HIGH"
},
"impactScore": 3.6,
"exploitabilityScore": 1.2
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Medtronic",
"product": "2090 CareLink Programmer",
"versions": [
{
"status": "affected",
"version": "All versions"
}
],
"defaultStatus": "unaffected"
},
{
"vendor": "Medtronic",
"product": "29901 Encore Programmer",
"versions": [
{
"status": "affected",
"version": "All versions"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2018-05-04T18:29:00.570",
"references": [
{
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-23"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-22"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system."
},
{
"lang": "es",
"value": "Todas las versiones de Medtronic 2090 Carelink Programmer se han visto afectadas por una vulnerabilidad de salto de directorio en la que la red de despliegue del software del producto podría permitir que un atacante lea archivos en el sistema."
}
],
"lastModified": "2026-06-17T02:00:20.220",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "884B08DC-B6F9-4B34-9679-61C256437DF9"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "215ED381-B433-4D05-97AD-4E7287E5820D"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}