CVE-2018-4846
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products).
Leer descripción completaMostrar menos
A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
CVSS
- Versión: 3.0
- Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 9.8
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 1.76%
- Percentil entre todas las CVEs puntuadas: 77
- Fecha de la puntuación: 5/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (3)
CWE
- CWE-798
- CWE-798
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2018-4846",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 10,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
"authentication": "NONE",
"integrityImpact": "COMPLETE",
"accessComplexity": "LOW",
"availabilityImpact": "COMPLETE",
"confidentialityImpact": "COMPLETE"
},
"acInsufInfo": false,
"impactScore": 10,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 10,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV30": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.0",
"baseScore": 9.8,
"attackVector": "NETWORK",
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "productcert@siemens.com",
"affectedData": [
{
"vendor": "Siemens AG",
"product": "RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems",
"versions": [
{
"status": "affected",
"version": "RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions >= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =< V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products"
}
]
}
]
}
],
"published": "2018-06-26T18:29:00.870",
"references": [
{
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
"tags": [
"Third Party Advisory"
],
"source": "productcert@siemens.com"
},
{
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
"tags": [
"Third Party Advisory"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "productcert@siemens.com",
"description": [
{
"lang": "en",
"value": "CWE-798"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-798"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
},
{
"lang": "es",
"value": "Se ha identificado una vulnerabilidad en los sistemas 1200 de RAPIDLab, los sistemas 400 de RAPIDPoint, los sistemas 500 de RAPIDPoint (todas las versiones sin usar productos Siemens Healthineers Informatics), las series 1200 de RAPIDLab (todas las versiones anteriores a la V3.3 con productos Siemens Healthineers Informatics), los sistemas 500 de RAPIDPoint (versiones 3.0 y posteriores con productos Siemens Healthineers Informatics), los sistemas 500 de RAPIDPoint (versiones 3.0 y posteriores con productos Siemens Healthineers Informatics), los sistemas 500 de RAPIDPoint (versiones 2.4.X con productos Siemens Healthineers Informatics) y los sistemas 400 de RAPIDPoint (todas las versiones con productos Siemens Healthineers Informatics). Una cuenta de fábrica con contraseña embebida podría permitir que los atacantes accedan al dispositivo por el puerto 5900/tcp. La explotación con éxito no requiere de interacción o privilegios de usuario e impacta en la confidencialidad, integridad y disponibilidad del dispositivo afectado. En el momento de la publicación del advisory, no se conoce ninguna explotación pública de la vulnerabilidad de seguridad. Siemens Healthineers confirma la vulnerabilidad de seguridad y proporciona mitigaciones para resolver el problema de seguridad."
}
],
"lastModified": "2026-06-17T01:59:09.657",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:siemens:rapidpoint_400_firmware:-:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "D3BD8C53-0B6B-4961-AF6E-302923B34118"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:siemens:rapidpoint_400:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "6081ED98-FD21-4C35-9E6F-95089EDB2905"
}
],
"operator": "OR"
}
],
"operator": "AND"
},
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:siemens:rapidpoint_500_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "15D5C974-4C15-450B-B9EF-51AE84A1741A",
"versionEndIncluding": "2.3"
},
{
"criteria": "cpe:2.3:o:siemens:rapidpoint_500_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "B2624D5C-08DF-4D18-8EC5-51509C2BFCE9",
"versionStartIncluding": "3.0"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:siemens:rapidpoint_500:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "9A4806F5-0292-45BA-96B9-422958D0CB0F"
}
],
"operator": "OR"
}
],
"operator": "AND"
},
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:siemens:rapidlab_1200_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "8A093C69-118A-44F9-98ED-B2B103B1EFBB",
"versionEndExcluding": "3.3"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:siemens:rapidlab_1200:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "757CBCB7-BE66-43E2-BD7D-55B67A649EBC"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "productcert@siemens.com"
}