« Volver al listado

CVE-2015-3954

Estado: ModificadaCrítica (9.8)—

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS

Probabilidad de explotación (EPSS)

EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).

Tecnologías afectadas (3)

CWE

Referencias

JSON original (NVD)

Mostrar
{
  "id": "CVE-2015-3954",
  "cveTags": [],
  "metrics": {
    "cvssMetricV2": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "version": "2.0",
          "baseScore": 10,
          "accessVector": "NETWORK",
          "vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
          "authentication": "NONE",
          "integrityImpact": "COMPLETE",
          "accessComplexity": "LOW",
          "availabilityImpact": "COMPLETE",
          "confidentialityImpact": "COMPLETE"
        },
        "acInsufInfo": false,
        "impactScore": 10,
        "baseSeverity": "HIGH",
        "obtainAllPrivilege": false,
        "exploitabilityScore": 10,
        "obtainUserPrivilege": false,
        "obtainOtherPrivilege": false,
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "type": "Primary",
        "source": "nvd@nist.gov",
        "cvssData": {
          "scope": "UNCHANGED",
          "version": "3.0",
          "baseScore": 9.8,
          "attackVector": "NETWORK",
          "baseSeverity": "CRITICAL",
          "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "integrityImpact": "HIGH",
          "userInteraction": "NONE",
          "attackComplexity": "LOW",
          "availabilityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "confidentialityImpact": "HIGH"
        },
        "impactScore": 5.9,
        "exploitabilityScore": 3.9
      }
    ]
  },
  "affected": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "affectedData": [
        {
          "vendor": "Hospira",
          "product": "Plum A+ Infusion System",
          "versions": [
            {
              "status": "affected",
              "version": "<= 13.4"
            }
          ]
        },
        {
          "vendor": "Hospira",
          "product": "Plum A+3 Infusion System",
          "versions": [
            {
              "status": "affected",
              "version": "<= 13.6"
            }
          ]
        },
        {
          "vendor": "Hospira",
          "product": "Symbiq Infusion System",
          "versions": [
            {
              "status": "affected",
              "version": "<= 3.13"
            }
          ]
        }
      ]
    }
  ],
  "published": "2019-03-25T17:29:00.670",
  "references": [
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "ics-cert@hq.dhs.gov"
    },
    {
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "source": "af854a3a-2127-422b-91ae-364da2661108"
    }
  ],
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "type": "Secondary",
      "source": "ics-cert@hq.dhs.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-285"
        }
      ]
    },
    {
      "type": "Primary",
      "source": "nvd@nist.gov",
      "description": [
        {
          "lang": "en",
          "value": "CWE-285"
        }
      ]
    }
  ],
  "descriptions": [
    {
      "lang": "en",
      "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
    },
    {
      "lang": "es",
      "value": "Hospira Plum A+ Infusion System versión 13.4 y anteriores, Plum A+3 Infusion System en su versión 13.6 y anteriores y Symbiq Infusion System, en su versión 3.13 y anteriores proporcionan privilegios root a usuarios no autenticados en el Puerto 23/TELNET por defecto. Un usuario no autorizado podría enviar comandos a la bomba. Hospira recomienda con los usuarios cierren el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado también el Plum 360 Infusion System que no es vulnerable a este problema."
    }
  ],
  "lastModified": "2026-06-17T00:26:32.110",
  "configurations": [
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:plum_a\\+_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2",
              "versionEndIncluding": "13.4"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:plum_a\\+_infusion_system:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "1DB2490B-0318-4770-BF45-CD7527F15D7F"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:plum_a\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "8536E705-89E8-47CB-9567-6AD65FBA0F1B",
              "versionEndIncluding": "13.6"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:plum_a\\+3_infusion_system:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "423AA561-8E38-4378-814B-1008B96F27A6"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "vulnerable": true,
              "matchCriteriaId": "C6DD5141-72AB-4694-8595-B4BED8EC7773",
              "versionEndIncluding": "3.13"
            }
          ],
          "operator": "OR"
        },
        {
          "negate": false,
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*",
              "vulnerable": false,
              "matchCriteriaId": "5C295E1A-BF60-476D-B972-5C5C28D7633B"
            }
          ],
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov"
}