CVE-2014-5433
Estado: ModificadaCrítica (9.8)—
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
CVSS
- Versión: 3.0
- Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Puntuación base: 9.8
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 2.06%
- Percentil entre todas las CVEs puntuadas: 81
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-312
- CWE-255
Referencias
JSON original (NVD)
Mostrar
{
"id": "CVE-2014-5433",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 7.5,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:L/Au:N/C:P/I:P/A:P",
"authentication": "NONE",
"integrityImpact": "PARTIAL",
"accessComplexity": "LOW",
"availabilityImpact": "PARTIAL",
"confidentialityImpact": "PARTIAL"
},
"acInsufInfo": false,
"impactScore": 6.4,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 10,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
],
"cvssMetricV30": [
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"scope": "UNCHANGED",
"version": "3.0",
"baseScore": 9.8,
"attackVector": "NETWORK",
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"integrityImpact": "HIGH",
"userInteraction": "NONE",
"attackComplexity": "LOW",
"availabilityImpact": "HIGH",
"privilegesRequired": "NONE",
"confidentialityImpact": "HIGH"
},
"impactScore": 5.9,
"exploitabilityScore": 3.9
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Baxter",
"product": "SIGMA Spectrum Infusion System",
"versions": [
{
"status": "affected",
"version": "6.05 (model 35700BAX) with wireless battery module (WBM) version 16"
}
]
}
]
}
],
"published": "2019-03-26T16:29:00.337",
"references": [
{
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-312"
}
]
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-255"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes."
},
{
"lang": "es",
"value": "Un atacante remoto no autenticado podría ser capaz de ejecutar comandos para visualizar credenciales de cuenta inalámbrica almacenadas en texto claro en Baxter SIGMA Spectrum Infusion System 6.05 (modelo 35700BAX), con un módulo de batería inalámbrica (WBM), en su versión 16, lo que podría permitir que un atacante obtenga acceso a la red host. Baxter ha publicado una nueva versión de SIGMA Spectrum Infusion System, la 8, que incluye cambios en el software y el hardware."
}
],
"lastModified": "2026-06-17T00:11:33.043",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "E6B02D87-6C75-4D60-8D4C-84628757214F"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "49E25260-EC14-4E98-A86B-CBBE47E26AE5"
},
{
"criteria": "cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "175BDC61-5163-4D78-BF02-6B9EF5D38841"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}