CVE-2014-5406
Estado: ModificadaAlta (9.3)—
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
CVSS
- Versión: 2.0
- Vector: AV:N/AC:M/Au:N/C:C/I:C/A:C
- Puntuación base: 9.3
Probabilidad de explotación (EPSS)
- Probabilidad de explotación en los próximos 30 días: 1.24%
- Percentil entre todas las CVEs puntuadas: 68
- Fecha de la puntuación: 6/10/2026
EPSS (Exploit Prediction Scoring System, de FIRST) estima la probabilidad de que una vulnerabilidad sea explotada en 30 días. Complementa a CVSS (impacto) y a CISA KEV (explotación confirmada).
Tecnologías afectadas (1)
CWE
- CWE-345
- CWE-345
Referencias
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
JSON original (NVD)
Mostrar
{
"id": "CVE-2014-5406",
"cveTags": [],
"metrics": {
"cvssMetricV2": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"cvssData": {
"version": "2.0",
"baseScore": 7.6,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:H/Au:N/C:C/I:C/A:C",
"authentication": "NONE",
"integrityImpact": "COMPLETE",
"accessComplexity": "HIGH",
"availabilityImpact": "COMPLETE",
"confidentialityImpact": "COMPLETE"
},
"acInsufInfo": false,
"impactScore": 10,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 4.9,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
},
{
"type": "Primary",
"source": "nvd@nist.gov",
"cvssData": {
"version": "2.0",
"baseScore": 9.3,
"accessVector": "NETWORK",
"vectorString": "AV:N/AC:M/Au:N/C:C/I:C/A:C",
"authentication": "NONE",
"integrityImpact": "COMPLETE",
"accessComplexity": "MEDIUM",
"availabilityImpact": "COMPLETE",
"confidentialityImpact": "COMPLETE"
},
"acInsufInfo": false,
"impactScore": 10,
"baseSeverity": "HIGH",
"obtainAllPrivilege": false,
"exploitabilityScore": 8.6,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
]
},
"affected": [
{
"source": "ics-cert@hq.dhs.gov",
"affectedData": [
{
"vendor": "Hospira",
"product": "LifeCare PCA Infusion System",
"versions": [
{
"status": "affected",
"version": "0",
"versionType": "custom",
"lessThanOrEqual": "5.0"
},
{
"status": "unaffected",
"version": "7.0"
}
],
"defaultStatus": "unaffected"
}
]
}
],
"published": "2015-07-06T19:59:00.097",
"references": [
{
"url": "http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/",
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
],
"vulnStatus": "Modified",
"weaknesses": [
{
"type": "Secondary",
"source": "ics-cert@hq.dhs.gov",
"description": [
{
"lang": "en",
"value": "CWE-345"
}
]
},
{
"type": "Secondary",
"source": "nvd@nist.gov",
"description": [
{
"lang": "en",
"value": "CWE-345"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459."
},
{
"lang": "es",
"value": "Hospira LifeCare PCA Infusion System anterior a 7.0 no valida trafico de la red asociado con el envío de (1) una librería de drogas, (2) una actualización de software o (3) un cambio de configuración, lo que permite a atacantes remotos modificar configuraciones o datos de medicamientos a través de paquetes en el puerto (a) TELNET, (b) HTTP, (c) HTTPS, o (d) UPNP. NOTA: este problema podría solapar el CVE-2015-3459."
}
],
"lastModified": "2026-06-17T00:11:30.157",
"configurations": [
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "6EF8A9BC-9349-4A57-A7B7-63640A066189",
"versionEndIncluding": "5.0"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:hospira:lifecare_pca3:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "86AC1C1C-D9FC-4EEE-B1A6-CEB03351EA58"
},
{
"criteria": "cpe:2.3:h:hospira:lifecare_pca5:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "C3551213-7D88-43AC-B56A-50F063884258"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov"
}